Acros e4401b service manual

Bike Brand: 
Bike Category: 
Road
e4401b service manual
LINK 1 ENTER SITE >>> Download PDF
LINK 2 ENTER SITE >>> Download PDF


File Name:e4401b service manual.pdf
Size: 2509 KB
Type: PDF, ePub, eBook
Category: Book
Uploaded: 5 May 2019, 20:14 PM
Rating: 4.6/5 from 766 votes.

Status: AVAILABLE


Last checked: 5 Minutes ago!

In order to read or download e4401b service manual ebook, you need to create a FREE account.

Download Now!

eBook includes PDF, ePub and Kindle version



✔ Register a free 1 month Trial Account.
✔ Download as many books as you like (Personal use)
✔ Cancel the membership at any time if not satisfied.
✔ Join Over 80000 Happy Readers


e4401b service manual

Please do not offer the downloaded file for sell only use it for personal usage. Looking for other manual. Document preview No preview for this item!For this no need registration. May be help you to repair. You could suffer a fatal electrical shock. Instead, contact your nearest service center. Note! To open downloaded files you need acrobat reader or similar pdf reader program. In addition, Also some files are djvu so you need djvu viewer to open them. These free programs can be found on this page: needed progs If you use opera you have to disable opera turbo function to download file. If you cannot download this file, try it with CHROME or FIREFOX browser. Translate this page: Relevant METER forum topics: HP8594 spektrumanalizator kalibracios problema Sziasztok Sajnos a fektett HP8594 spektrumanalizator megadta magat. Bekapcsolaskor kiirja a kepernyore hogy FREQ UNCAL A CAL OUT kimenetet osszekotom az RF Input-al, es nyomok egy kalibraciot. Kiirja a kepernyore hogy CAL SIGNAL NOT FOUND Keresgeltem az internetet es irjak hogy a belso elem lemerult, ezert kalibracios adatokat elfelejtette. De a datumot es az idot tovabbra is tarolja. Illetve miert irja kalibraciokor hogy CAL SIGNAL NOT FOUND Mellekelem a fotokat. Esetleg talalkozott ilyennel. Koszi a valaszokat elore is Velleman HPS140 pocket scope (segitseg kellene) Udv Tanyalakok,Epito jellegu valaszaitokat elore is koszonom.A HP 54503 oszcilloszkop kovetkezo hibaval uzemkeptelen. A menu rendesen mukodik. Tapegyseg rendesen kiadja a feszultseget. Lehetseges hiba a fuggoleges elteritesnel a RAM a processzor korul. Kapcsolasi rajzom nincs, Nem tudom honnan lehet letolteni. Kosuzonom a hozzaszolasokat. A muszer megjavitva. Kulon koszonet, akik utaltak a hibas NV RAM aalkatreszre, es a benne elhelyezkedo gombelemre. Ez volt a hibas alkatresz. A DALLAS alkatresz belsejeben egy gombelemre kiszerelve, es hehere egy masik gombelemre cserelve. Utamna az osszes selfteszt passos, a kalibracio sikeres.

http://giorgerika.com/craftsman-6_5-lawn-mower-manual.xml

    Tags:
  • e4401b service manual, e4401b service manual, e4407b service manual, e4401b service manuals, e4402b service manual.

Azota nem ir ki hibat, a szkop tokeletesen mukodik. Udv. Esetleg letezik egy mester kod, ill valamilyen modszer a kiolvasasra. Amihez nem kell GB-IB vagy HP-IB kabelt venni. A muszer tokeletes, pontos, csak zavar, hogy le van zarva, igaz igy bolondbiztos, es ugyis lezarnam elnyomkodas ellen, aztan elfelejtenem az uj kodot, de en felmatricaznam belul. Tenyleg lehet szet kellene szednem, hatha belul van egy 12 digites szamsor.Majd holnap. Similar manuals: You can write in English language into the forum (not only in Hungarian). Shop for Comparable Products on our Paste together up to 32 discontinuous frequency or zero spans in one sweep. Eliminate the need for sweeping through unwanted frequencies and for multiple setups. Variable sweep (trace) points, ranging from 101 to 8192 points, optimize measurements for frequency resolution and accuracy versus speed. 16.8 cm, high-resolution color display with wide viewing angle make it easy to identify signals of interest. Split screen display shows wide spans while zooming in on signals of interest. The next generation user interface improves ease-of-use. Built-in help gives immediate assistance without manuals. One-button measurements such as ACP, Occupied BW, Emissions BW, channel power, 10 peaks table and harmonic distortion provide faster and more repeatable results. Built-in frequency counter with 1 Hz resolution precisely measures individual signals. Specifications. Frequency Range: 9 kHz to 1.5 GHz. Frequency Span Range: 0 Hz (zero span), 100 Hz to the range of the spectrum analyzer. Amplitude Measurement Range: Displayed average noise level to maximum safe input level. AYX Fast zero-span sweeps. B72 Increases useable memory to 10 MB. BAA FM Demodul Pricing does include NIST traceable calibration and issue of a calibration certificate and calibration label. Pricing may vary slightly due to volume and location of laboratory supporting calibration. Volume pricing may apply.

http://auto-rujo.com/images-editor/craftsman-6_75-hp-21-lawn-mower-manual.xml

On-site fees may apply depending on logistics, location and volume of work to be completed during the visit. For a complete list go to our Modulation Modulation is the coding of information onto the carrier frequency. This includes amplitude, frequency, or phase modulation techniques. Sensitivity Sensitivity is the smallest signal that can reliably detected. Sensitivity specifies the strength of the smallest signal at the input that causes the output signal power to be M times the output noise power where M must be specified. ASC ATCI ATM Microwave AVL CPI Cross Technologies Datum Dielectric - SPX Divicom Drake Microwave Systems. Midas Mini-Circuits Mitec Miteq Mobile Pathways Mu-Del Electronics Narda New and Used inc. New Era Systems New Japan Radio Newtec Norsat Satellite Transmission Systems SeaTel Sector Microwave Sector Motor Sew-Eurodrive SKYWARE GLOBAL Smith Communications SMT Solasat Sony Specialty Microwave Specialty Microwave Corp Standard Communications Storm Flex Struthers Electornic Sumitomo Superior Electric Swedish Microwave Symmetricom We now have an online store to quickly and easily order any accessories you may need. This manual provides documentation for the following instruments: Agilent Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Safety Notes This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.

http://gbb.global/blog/bose-v20-owners-manual

The following examples illustrate warning and caution statements that are used throughout this manual. WARNING Warning denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in injury or loss of life. Do not proceed beyond a warning note until the indicated conditions are fully understood and met. CAUTION Caution denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of the instrument. Do not proceed beyond a caution sign until the indicated conditions are fully understood and met. The following safety notes are used throughout this manual. Familiarize yourself with each of the notes and its meaning before operating this instrument. WARNING This is a Safety Class 1 Product (provided with a protective earth ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited. 2 3 WARNING These servicing instructions are for use by qualified personnel only. To avoid electrical shock, do not perform any servicing unless you are qualified to do so. WARNING If this product is not used as specified, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only. CAUTION This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 101 and 664 respectively. Warranty This Agilent Technologies instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Agilent Technologies will, at its option, either repair or replace products which prove to be defective.

http://8forwine.com/images/canon-powershot-a720-is-manual.pdf

For warranty service or repair, this product must be returned to a service facility designated by Agilent Technologies. Buyer shall prepay shipping charges to Agilent Technologies and Agilent Technologies shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to Agilent Technologies from another country. Agilent Technologies warrants that its software and firmware designated by Agilent Technologies for use with an instrument will execute its programming instructions when properly installed on that instrument. Agilent Technologies does not warrant that the operation of the instrument, or software, or firmware will be uninterrupted or error-free. 3 4 LIMITATION OF WARRANTY The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.Refer to Parts List in Chapter 5. DO NOT THROW BATTERIES AWAY BUT COLLECT AS SMALL CHEMICAL WASTE. It provides procedures for troubleshooting common failures and isolating problems in the analyzer. The analyzer contains static sensitive components. Read the section entitled ESD Information on page 1-15 in this chapter. Become familiar with the organization of the troubleshooting information in this chapter. Read the rest of this section. WARNING These servicing instructions are for use by qualified personnel only. WARNING The opening of covers or removal of parts is likely to expose dangerous voltages. Disconnect the product from all voltage sources while it is being opened. WARNING The detachable power cord is the instrument disconnecting device. It disconnects the mains circuits from the mains supply before other parts of the instrument.

The front panel switch is only a standby switch and is not a LINE switch (disconnecting device). CAUTION Always position the instrument for easy access to the disconnecting device (detachable power cord). WARNING To prevent electrical shock, disconnect the analyzer from mains before cleaning. Use a dry cloth or one slightly dampened with water to clean the external case parts. Do not attempt to clean internally. WARNING For continued protection against fire hazard, replace fuses, and or circuit breakers only with same type and ratings. The use of other fuses, circuit breakers or materials is prohibited. Chapter 1 13 14 Troubleshooting the Analyzer Before You Start WARNING This is a Safety Class 1 Product (provided with a protective earth ground incorporated in the power cord). Intentional interruption is prohibited. CAUTION Always use the three-prong ac power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage. CAUTION Before switching on this instrument, make sure of the following: the correct fuse is installed the supply voltage is in the specified range CAUTION This instrument has auto-ranging line voltage input, be sure the supply voltage is within the specified range and voltage fluctuations do not to exceed 10 percent of the nominal supply voltage. WARNING The Agilent Technologies E4402B, E4403B, E4404B, E4405B, E4407B, E4408B, E4411B, E7402A, E7405A may be used with an E1779B Rechargeable Battery Pack which contains a Ni-Cd battery. The battery must be recycled or disposed of properly.The following assemblies must be replaced as an assembly. After an Analyzer Repair If one or more analyzer assemblies have been repaired or replaced, perform the related adjustments and performance verification tests. Refer to Table 7-2 in Chapter 7 for the related adjustments and performance verification tests required for each assembly.

ESD Information Protection from Electrostatic Discharge Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Figure 1-1 shows an example of a static-safe workstation using two types of ESD protection: o o Conductive table-mat and wrist-strap combination. Conductive floor-mat and heel-strap combination. Chapter 1 15 16 Troubleshooting the Analyzer Before You Start Both types, when used together, provide a significant level of ESD protection. Of the two, only the table-mat and wrist-strap combination provides adequate ESD protection when used alone. To ensure user safety, the static-safe accessories must provide at least 1 megohm of isolation from ground. Refer to Table 1-1 on page 16 for information on ordering static-safe accessories. WARNING Figure 1-1 These techniques for a static-safe workstation should not be used when working on circuitry with a voltage potential greater than 500 volts. The risk of ESD damage can be greatly reduced by close attention to how all components are handled.Keep static-generating materials at least one meter away from all components. Store or transport components in static-shielding containers. CAUTION Always handle printed circuit board assemblies by the edges. This will reduce the possibility of ESD damage to components and prevent contamination of exposed plating. Test Equipment Usage and ESD o o o Before connecting any coaxial cable to an analyzer connector, momentarily short the center and outer conductors of the cable together. Personnel should be grounded with a 1 megohm resistor-isolated wrist-strap before touching the center pin of any connector and before removing any assembly from the analyzer. Be sure that all analyzers are properly earth-grounded to prevent build-up of static charge. The ESD standards developed by this agency are sanctioned by the American National Standards Institute (ANSI).

18 Chapter 1 19 Troubleshooting the Analyzer Check the Basics Check the Basics A problem can often be resolved by repeating the procedure you were following when the problem occurred. Before calling Agilent Technologies or returning the analyzer for service, please make the following checks: o o o o o o o Check the line fuse. Is there power at the receptacle. Is the analyzer turned on. Make sure the fan is running, which indicates that the power supply is on. If the display is dark or dim, press the upper Viewing Angle key in the upper-left corner of the front panel. If the display is too bright, adjust the lower Viewing Angle key in the upper-left corner of the front panel. If other equipment, cables, and connectors are being used with your analyzer, make sure they are connected properly and operating correctly. Review the procedure for the measurement being performed when the problem appeared. Are all the settings correct. If the analyzer is not functioning as expected, return the analyzer to a known state by pressing the Preset key and Factory Preset (if present). Some analyzer settings are not affected by a Preset. Refer to the Front-Panel Key Reference chapter of the user s guide for information on settings affected by a Preset.Refer to the appropriate Specifications and Characteristics chapters in the specifications guide for more details. In order to meet specifications, the analyzer must be aligned. Either Auto Align All must be selected (press System, Alignments, Auto Align, All), or the analyzer must be manually aligned at least once per hour, or whenever the temperature changes more than 3 C. When Auto Align, All is selected, AA appears on the left edge of the display. Refer to the appropriate Specifications and Characteristics chapters in the specifications guide for more details. Is the analyzer displaying an error message.

If the necessary test equipment is available, perform the performance verification tests in the calibration guide or use the performance verification software. Record all results on a Performance Verification Test Record form which follows the tests. If the equipment to perform the performance verification tests is not available, you may still be able to perform the functional checks in the user s guide. Chapter 1 19 20 Troubleshooting the Analyzer Problems at Instrument Power-Up Problems at Instrument Power-Up This section describes symptoms that can occur when the analyzer is first powered on. CAUTION Immediately unplug the analyzer from the ac power line if the unit shows any of the following symptoms: Smoke, arcing, or unusual noise from inside the unit, except as noted below. No response of any kind when unit is plugged into ac power mains and turned on. The analyzer ac power fuse blows. A circuit breaker or fuse on the main ac power line opens. These potentially serious faults must be corrected before proceeding. Refer to Troubleshooting an Inoperative Analyzer on page 21. Note that these analyzers are equipped with a speaker and normally emit noise from the speaker at power-up. This is not a problem; adjust the volume control as desired. The analyzer will also occasionally emit clicking noises. These are normal and are related to the analyzer self alignment routines. If the analyzer always turns on when power is applied, check the rear of the analyzer. There is an always on power mode switch that can be changed. Refer to the rear panel features section of the user s guide for information on this switch. 20 Chapter 1 21 Troubleshooting the Analyzer Troubleshooting an Inoperative Analyzer Troubleshooting an Inoperative Analyzer When an analyzer appears to be dead (no display or inoperative fan) there is often little evidence that points directly to the cause. This section provides steps and solutions to typical failure modes relating to an inoperative analyzer.

Before troubleshooting an analyzer, ensure that it has been set up correctly. Check the Analyzer Setup Perform the steps in the section titled Check the Basics earlier in this chapter. Initial Checks Perform the following initial checks when first troubleshooting an inoperative analyzer. 1. Is the fan running? a. If the fan is not running, refer to the section If the Fan is Not Operating on page 22. b. If the fan is running, this indicates the power supply is functioning at least partially. 2. Is the display on? a. If the display is blank, refer to the section Checking the Power Supplies on page 23. Also verify that the firmware is good by performing the Bootrom Self-Test Check on page 38. b. If the display is on and the fan is running, suspect one or more power supplies is being pulled down, or a defective processor assembly. In either case, the analyzer outer case and inner shield must be removed. Refer to Chapter 6 for removal procedures. 3. Check the power supplies. a. To check the power supply voltages, refer to the section Checking the Power Supplies on page 23. b. If the supplies all measure good, suspect a defective processor, or a defective LCD or interface board. Chapter 1 21 22 Troubleshooting the Analyzer Troubleshooting an Inoperative Analyzer 4. Perform the following checks for the processor assembly. a. Is the analyzer clicking. The analyzer should occasionally emit clicking noises. These are related to the analyzer automatic self alignment routines, and indicate the processor is basically functioning. When the analyzer is cold, there will be more self alignment activity (more clicking) than when it is warm. If you do not hear any clicks after a few minutes, suspect a defective processor assembly. Refer to the section Troubleshooting the A4 Processor Assembly on page 37. b. Are the processor assembly LEDs all out. There are LEDs along the top edge of the processor assembly. These should all be off.

To check their functionality, cycle the analyzer power. The processor LEDs should all go on for a few seconds, then go off. If one or more LEDs remains on, suspect the processor assembly is defective. Refer to the section Bootrom Self-Test Check on page 38. c. If the analyzer is clicking, and all of the processor LEDs are out, suspect a defective LCD or a defective front panel interface board. Refer to the section Troubleshooting the LCD Display on page 25. If the Line Fuse Has Blown If the line fuse has blown, perhaps a nonstandard fuse with too low a current rating was installed. If the line fuse still blows, suspect that the power supply assembly is defective. If the Fan is Not Operating CAUTION The power supply may be hot if the instrument has been operating without the fan running. Allow the instrument to cool down before troubleshooting. 1. If there is no display: a. Unplug the line-power cord. b. Change the switch at the rear of the analyzer so the power is always on. Refer to the rear panel features section of the user s guide for information on this switch. This will bypass the front panel power switch. 22 Chapter 1 23 Troubleshooting the Analyzer Troubleshooting an Inoperative Analyzer c. If the fan still does not turn on, suspect a defective power supply assembly. Refer to the section Checking the Power Supplies on page 23, to check individual supply voltages. d. If the fan starts, this indicates a problem with the front panel switch or the circuitry to the front panel switch. 2. If there is a display: a. Remove the analyzer outer case. Refer to Chapter 6 for removal details. b. Measure the fan voltage. c. The fan voltage can be checked directly at the center of the fan where the wires connect. The fan normally operates on a voltage ranging from 5 to 16 Vdc depending on the temperature of the analyzer. As analyzer temperature increases, fan voltage should become more negative, increasing the speed of the fan.

If the correct voltage is present, suspect a defective fan. If the voltage is not present, check the plug where the fan wires attach to the power supply board. If this looks good, suspect a defective power supply. Refer to Chapter 6 for information on how to remove the power supply. Checking the Power Supplies The power supply test board is no longer supplied. If the power supply is suspect, change the power supply. All ESA models use the same power supply, so substitution is possible to determine if the power supply is defective. WARNING The analyzer contains potentially hazardous voltages. Refer to the safety symbols provided on the analyzer, and in the general safety instructions in this guide, before operating the unit with the cover removed. Ensure that safety instructions are strictly followed. Failure to do so can result in severe or fatal injury. See Figure 1-2 for these test points. If 12.6 V is within limits, it is a reasonably good indicator that 15 V is within specifications. Chapter 1 23 24 Troubleshooting the Analyzer Troubleshooting an Inoperative Analyzer Figure 1-2 Probe Power Connector Voltages If All Low Voltage Supplies are At or Near Zero Volts If all the power supplies are dead, suspect a defective A5 power supply assembly. Refer to Chapter 6 for removal instructions. If Some Low Voltage Supplies are At or Near Zero Volts If some supplies are functioning while others are dead or low, it is possible that one of the assemblies in the analyzer is loading down the power supply. In this case it is necessary to sequentially remove the assemblies, taking care to disconnect the line-power cord before removing any assembly. A logical first step would be to unplug the RF assembly ribbon cable from the motherboard at the rear of the instrument. This disconnects the complete RF section without having to remove it from the chassis. Refer to Chapter 6 for any other assembly removal procedures.

After an assembly is disconnected or removed, plug the line-power cord back into the analyzer and remeasure the supply that was down. If it is still down, continue with the assembly removal. If the supply is now up, suspect the last assembly removed as being defective. 24 Chapter 1 25 Troubleshooting the Analyzer Troubleshooting the LCD Display Troubleshooting the LCD Display The only adjustment that can be made to the LCD display is the viewing angle. This is found on the front panel of the analyzer in the upper left corner. Viewing Angle 1. Verify that the analyzer has gone through a full power-on sequence. Refer to the section Troubleshooting an Inoperative Analyzer on page If the display is dark (not visible), connect an external VGA monitor to the rear panel VGA output connector on the analyzer. It is possible that some multisync monitors may not be able to lock to a 60 Hz sync pulse. If the video information is not present on the external VGA monitor, the most probable cause is the A4 processor assembly. 3. If the external VGA monitor is functioning, verify that the ribbon cables going to and from the A1A1 front panel interface assembly are aligned properly and securely plugged into the connectors. 4. To determine whether the A1A2 LCD display or the A1A1 front panel interface assembly is defective, remove the front frame assembly. WARNING Table 1-2 The inverter board contains high voltage for the backlights. Carefully remove the inverter board by taking out the two screws securing it to the front frame, then turn it over to access the CN1 connector. Refer to Table 1-2 on page 25 and carefully measure the voltage coming into the inverter board at CN1 pins 1-6 with a digital voltmeter. Measurement Locations and Expected Voltages Measurement Location CN1 Pin 1 CN1 Pin 2 CN1 Pin 3 CN1 Pin 4 CN1 Pin 5 CN1 Pin 6 Expected Voltage from A1A1 Front Panel Interface 0 Vdc 0 Vdc 5 Vdc 5 Vdc 30 mv 2.

5 Vdc The display is not field-repairable, and must be replaced as an assembly. Refer to Chapter 5 for part-number information. Chapter 1 25 26 Troubleshooting the Analyzer Troubleshooting the LCD Display Figure 1-3 LCD Troubleshooting Measurement Locations If the voltages are correct, the most probable cause of failure is the A1A2 LCD assembly. NOTE The LCD has two backlights to illuminate the LCD. If both backlights are defective, the screen will be dark. The backlights can be replaced individually, however, it is improbable that both backlights are defective simultaneously. If one backlight is defective, the LCD will illuminate but the dimness will be noticeable, and you will need to replace A1A2DS1 or A1A2DS2. It is recommended that both backlights be replaced at the same time if it is likely that one simply wore out. 26 Chapter 1 27 Troubleshooting the Analyzer Isolating an LO, IF, or Video Problem Isolating an LO, IF, or Video Problem This section provides techniques for isolating amplitude failures along the signal path. Checking Detector Modes The instrument has three detector modes known as peak, negative peak, and sample. All of these detectors are located on the A4 processor board. Normal instrument power up routes the detected video signal through the positive peak detector only. If there are no graticules, signal information, or screen annotations, refer to the section Troubleshooting the LCD Display on page 25. The 2 Vdc can be measured on the motherboard at A7J13 pin 24 if the analyzer is an E4403B, E4408B, or E4411B. If you do not measure a dc level at approximately 2 Vdc, the IF assembly is the most probable cause. Refer to Figure If you measure 2 Vdc on the output of the IF assembly but do not see a flat line at the top of the graticule, there may be a problem between the output of the IF assembly and the input to the A4 processor board.

Refer to section Using the Internal Service-Diagnostic Routines on page 32 for more information about the video shift function. 3. If the power hole or power spike does not go away after re-adjusting the frequency response (flatness), the RF assembly is the most probable cause. If the analyzer is in full span and the 21.4 MHz input frequency has a power level of 20 dbm, the displayed response will be the inverse of the RF assembly flatness. In narrow spans, the flatness correction will be done by changing the A3 IF assembly gain. In large spans, when it is not practical to change the A3 IF assembly gain as the analyzer sweeps, video shift is used for flatness correction. Otherwise: 1. Perform Align Now All by pressing System, Alignments, Align Now, All and wait for the alignment to complete. Three complete alignment sequences will be performed. 2. Turn Auto Align back on by pressing System, Alignments, Auto Align, All. 3. Verify the analyzer performance now that the A3 IF assembly has had the defaults loaded and the auto align has been turned back on. 28 Chapter 1 29 Troubleshooting the Analyzer Isolating an LO, IF, or Video Problem 4. If the signal on screen is still not correct perform Checking Detector Modes on page 27. Now inject a 21.4 MHz signal at 10 dbm from a synthesizer into the A3 IF assembly. With the reference level at 0 dbm, the input attenuator set to 0 db, and center frequency at 50 MHz in zero span, a flat line should appear at the top graticule. 5. If the signal is not at the top graticule, measure A7J13 pin 24 on the motherboard with a voltmeter and verify 2 Vdc. Refer to Figure If 2 Vdc is present the most probable cause is either an A7 motherboard trace problem to the A4 processor board, a bent pin on A7J7 motherboard connector, or the A4 processor board itself. Checking a Horizontal Trace With No Signal or Noise Floor Present 1. For E4402B, E4404B, E4405B, and E4407B connect the Amptd Ref out to the Input.

Bike Model Name: 
e4401b service manual